
UPC-A
Product Reference Guide 5-7
UPC-A — continued
Add-On Timer
This feature sets the period of time after a UPC-A label is decoded when
optional add-on reading is enabled.
START
STATE BAR CODE
This item is used to set the timer value when optional
add-ons are enabled. The selectable range is from
010 to 300 mSec in 10 mSec increments. To config-
ure this feature, scan the “START” bar code above to
place the unit in Programming Mode, then the “Add-
On Timer,” followed by the three digits (zero padded)
from the Alphanumeric table in
Appendix C, Alpha-
Numeric Pad
representing the disired timer setting.
Exit programming mode by scanning the “END” bar
code below.
Add-On Timer
DEFAULT SETTING FOR THIS FEATURE:
007 (70 milliseconds)
END
Komentarze do niniejszej Instrukcji