General Features
14
Magellan
TM
1100i
ERI Timeout
Specifies the amount of time the External Read Indicator (ERI) signal is held active for a
good read.
START / END
PROGRAMMING BARCODES
Sets the ERI timeout duration using hex
values from 000 to 255 in increments of ten milliseconds (10ms or
0.01 seconds). To configure this feature, scan the “START/END” barcode above to place the unit in Program-
ming Mode, then the “Set ERI Timeout,” followed by the two digits (zero padded) from the Alphanumeric
t
able in
Appendix C, Alpha-Numeric Pad representing the desired time value. Exit programming mode by scan-
ning the “START/END” barcode again.
DEFAULT SETTING FOR THIS FEA
TURE: 20 milliseconds (02)